EURL — European Union reference laboratory

A laboratory designated under the IVDR to verify the performance of class D devices and check batches, and to provide scientific and technical assistance to the Commission and Member States. There is no direct MDR equivalent.

Reference: IVDR Article 100

Read More

Expert panel

A body of independent experts appointed to give scientific, technical and clinical opinions, including views on clinical evaluation under the consultation procedure and on common specifications.

Reference: MDR Article 106

Read More
EU, EU MDR, EU IVDR, Device types EU, EU MDR, EU IVDR, Device types

Falsified device

Any device with a false presentation of its identity and/or of its source and/or its CE marking certificates or documents relating to CE marking procedures. This definition does not include unintentional non-compliance and is without prejudice to infringements of intellectual property rights.

Reference: MDR Article 2(9); IVDR Article 2(10)

Read More
EU, EU MDR, EU IVDR, Vigilance EU, EU MDR, EU IVDR, Vigilance

Field safety corrective action

Corrective action taken by a manufacturer for technical or medical reasons to prevent or reduce the risk of a serious incident in relation to a device made available on the market.

Reference: MDR Article 2(68); IVDR Article 2(71)

Read More
EU, EU MDR, EU IVDR, Vigilance EU, EU MDR, EU IVDR, Vigilance

Field safety notice

A communication sent by a manufacturer to users or customers in relation to a field safety corrective action.

Reference: MDR Article 2(69); IVDR Article 2(72)

Read More

Fully refurbishing

For the purposes of the definition of manufacturer, the complete rebuilding of a device already placed on the market or put into service, or the making of a new device from used devices, to bring it into conformity with this Regulation, combined with the assignment of a new lifetime to the refurbished device.

Reference: MDR Article 2(31); IVDR Article 2(24)

Read More
EU, EU MDR, EU IVDR, Transition EU, EU MDR, EU IVDR, Transition

Gap assessment

A structured comparison of existing technical documentation and processes against the requirements that now apply, producing a list of what must change. An industry practice rather than a regulatory requirement, most often run when moving a device from the directives to the regulations.

Read More
EU, EU MDR, EU IVDR, Device types EU, EU MDR, EU IVDR, Device types

Generic device group

A set of devices having the same or similar intended purposes or a commonality of technology allowing them to be classified in a generic manner not reflecting specific characteristics.

Reference: MDR Article 2(7); IVDR Article 2(8)

Read More
EU, EU MDR, Transition EU, EU MDR, Transition

Grandfathering

Carrying a device’s existing approval forward into a new regulatory framework without reassessment. The MDR contains no such mechanism: devices certified under the directives must be assessed against the regulation, with the transitional provisions governing only how long the old certificates remain valid.

Read More

GSPR — General safety and performance requirements

The set of requirements every device must meet before it can be CE marked, covering general safety, design and manufacture, and the information supplied with the device. Compliance is demonstrated requirement by requirement, and the evidence sits in the technical documentation. Requirements that do not apply to a given device must be justified rather than ignored.

Reference: MDR Annex I; IVDR Annex I

Read More

Harmonised standard

A European standard as defined in point (1)(c) of Article 2 of Regulation (EU) No 1025/2012.

Reference: MDR Article 2(70); IVDR Article 2(73)

Read More

Health institution

An organisation the primary purpose of which is the care or treatment of patients or the promotion of public health. Under the IVDR: an organisation the primary purpose of which is the care or treatment of patients or the promotion of public health; 02017R0746 — EN — 10.01.2025 — 005.001 — 7.

Reference: MDR Article 2(36); IVDR Article 2(29)

Read More

HIE — Health institution exemption

The route allowing a health institution to manufacture and use devices internally without full conformity assessment, subject to conditions: the devices are not transferred to another legal entity, they are made under an appropriate quality management system, a public declaration is drawn up, and no equivalent device on the market meets the patient group’s needs at the required performance level.

Reference: MDR Article 5(5); IVDR Article 5(5)

Read More
EU, EU MDR, EU IVDR, Traceability EU, EU MDR, EU IVDR, Traceability

HRI — Human readable interpretation

A legible interpretation of the data characters encoded in the UDI carrier.

Reference: MDR Annex VI Part C Section 1; IVDR Annex VI Part C Section 1

Read More

IMDRF — International Medical Device Regulators Forum

A voluntary group of medical device regulators working towards converged regulatory approaches. Its documents carry no legal force in the EU, but several are referenced by MDCG guidance and are used to fill gaps where EU guidance is silent.

Read More

Implantable device

Any device, including those that are partially or wholly absorbed, which is intended: — to be totally introduced into the human body, or — to replace an epithelial surface or the surface of the eye, by clinical intervention and which is intended to remain in place after the procedure. Any device intended to be partially introduced into the human body by clinical intervention and intended to remain in place after the procedure for at least 30 days shall also be deemed to be an implantable device.

Reference: MDR Article 2(5)

Read More

Implant card

The card supplied with an implantable device carrying the information needed to identify it, including name, serial or lot number and UDI, together with warnings and precautions. The health institution makes it available to the patient.

Reference: MDR Article 18

Read More

Importer

Any natural or legal person established within the Union that places a device from a third country on the Union market.

Reference: MDR Article 2(33); IVDR Article 2(26)

Read More
EU, EU IVDR, IVD, Device types EU, EU IVDR, IVD, Device types

In vitro diagnostic medical device

Any medical device which is a reagent, reagent product, calibrator, control material, kit, instrument, apparatus, piece of equipment, software or system, whether used alone or in combination, intended by the manufacturer to be used in vitro for the examination of specimens, including blood and tissue donations, derived from the human body, solely or principally for the purpose of providing information on one or more of the following: (a) concerning a physiological or pathological process or state; (b) concerning congenital physical or mental impairments; (c) concerning the predisposition to a medical condition or a disease; (1) Directive 2006/42/EC of the European Parliament and of the Council of 17 May 2006 on machinery (OJ L 157, 9.6.2006, p. 24). 02017R0746 — EN — 10.01.2025 — 005.001 — 4 (d) to determine the safety and compatibility with potential recipients; (e) to predict treatment response or reactions; (f) to define or monitoring therapeutic measures. Specimen receptacles shall also be deemed to be in vitro diagnostic medical devices.

Reference: IVDR Article 2(2)

Read More
EU, EU MDR, EU IVDR, Vigilance EU, EU MDR, EU IVDR, Vigilance

Incident

Any malfunction or deterioration in the characteristics or performance of a device made available on the market, including use-error due to ergonomic features, as well as any inadequacy in the information supplied by the manufacturer and any undesirable side-effect. Under the IVDR: any malfunction or deterioration in the characteristics or performance of a device made available on the market, including use-error due to ergonomic features, as well as any inadequacy in the information supplied by the manufacturer and any harm as a consequence of a medical decision, action taken or not taken on the basis of information or result(s) provided by the device.

Reference: MDR Article 2(64); IVDR Article 2(67)

Read More